Zurück
Vor
 
 
HI88703 Precision Turbidity Benchtop Meter, USEPA Compliant
HI88703 Precision Turbidity Benchtop Meter, USEPA Compliant
THB 80,000.00 * Prices plus VAT plus shipping costs
Star Rating (0)
 
 
Amount:
 
 
 
 

Product information "HI88703 Precision Turbidity Benchtop Meter, USEPA Compliant"

Two measuring ranges: Ratio turbidity, non-ratio turbidity

  • Meets USEPA requirements
  • GLP features        
  • Two, three, four or five point turbidity calibration
  • USB PC connectivity
  • Log up to 200 measurements
  • Contextual help and tutorial mode

 

The HI 88703 turbidity benchtop meter is specially designed for water quality measurements, providing reliable and accurate readings on low turbidity ranges.

This instrument has an EPA compliance reading mode which rounds the reading
to meet EPA reporting requirements. Alternative EBC and Nephelos measuring units are available. Depending on the measured sample and needed accuracy, normal, continuous or signal averaging measurement can be selected.

A two, three, four or five-point calibration could be performed by using the supplied standards. When user prepared standards are used, calibration points can be modified.
H 88703 features complete GLP (Good Laboratory Practice) functions that allows traceability of the calibration conditions. The last calibration points, time and date can be checked.

Up to 200 measurements can be stored in internal memory. Data can be transferred to a PC via optional HI 920013 USB cable and HI 92000 Windows® compatible software.

 

Order Information:
HI 88703-02 (230V) is supplied with sample cuvettes and caps (5), calibration cuvettes, silicone oil (HI 93703-58), tissue for wiping cuvettes, power cord and instruction manual.

Product Manuals

  Manual: Download

Specifications :

Non Ratio Mode 0.00 to 9.99; 10.0 to 40.0 NTU;
0.0 to 99.9; 100 to 268 Nephelos
0.00 to 9.80 EBC
  Ratio Mode 0.00 to 9.99; 10.0 to 99.9; 100 to 4000 NTU
0.0 to 99.9; 100 to 26800 Nephelos
0.00 to 9.99; 10.0 to 99.9; 100 to 980 EBC
Resolution Non Ratio Mode 0.01; 0.1 NTU 0.1; 1 Nephelos; 0.01 EBC
  Ratio Mode 0.01; 0.1; 1 NTU 0.1; 1 Nephelos 0.01; 0.1, 1 EBC
Range Selection   automatic
Accuracy @25°C   ±2% of reading plus 0.02 NTU (0.15 Nephelos; 0.01 EBC)
±5% of reading above 1000 NTU (6700 Nephelos; 245 EBC)
Repeatability   ±1% of reading or 0.02 NTU (0.15 Nephelos; 0.01 EBC)whichever is greater
Stray Light   < 0.02 NTU (0.15 Nephelos; 0.01 EBC)
Light Detector   silicon photocell
Method   Nephelometric method (90°) or Ratio Nephelometric Method(90° & 180°), Adaptation of the USEPA Method 108.1 and Standard Method 2130 B.
Measuring Mode   normal, average, continuous
Turbidity Standards   <0.1, 15, 100, 750 and 2000 NTU
Calibration   two, three, four or five-point calibration
Light Source/Life   tungsten filament lamp / greater than 100,000 readings
Display   40 x 70 mm graphic LCD (64 x 128 pixels) with backlight
Log Memory   200 records
PC Interface   USB
Environment   0°C (32°F) to 50°C (122°F); max 95% RH non-condensing
Power Supply   230 V/50 Hz or 115 V/60 Hz 20 W; auto-off after 15 minutes of non-use
Dimensions   230 x 200 x 145 mm (9 x 7.9 x 5.7”)
Weight   2.5 Kg (88 oz.)

 

 

More Information about "HI88703 Precision Turbidity Benchtop Meter, USEPA Compliant"